Basic Information
Product Name
Aed Defibrillator Monitor
Audible Prompts
22 Audible Prompts
Operation Temp
0°c to 50°c
Product Description
SY-C025 Hot Sale Hospital Medical Defibrillator Monitor AED
| Output | Biphasic Truncated Exponential |
| Energy Sequence | 150, 150, 200J |
| Charge Time | 8 sec. to 150J; 10 sec. to 200J (New Battery) |
| Analysis Time | 9 sec. |
| Max Time to Readiness (New Battery) | Less than 30 seconds |
| Max Time to Readiness (After 6 shocks) | Less than 35 seconds |
| Max Time from Power-on to Readiness | Less than 25 seconds |
| Audible Prompts | 22 audible prompts |
| Visual Prompts | LED prompts |
| Controls | Two buttons-ON/Off, Shock |
| Output Energy Accuracy | ±15% into any impedance from 25 to 175Ω |
| The Maximum Voltage | 1200±50V |
| Patient Impedance Limits | 20Ω to 200Ω |
Product Visuals
Operating Guide & Interface
Frequently Asked Questions
How long does it take for the SY-C025 to analyze a heart rhythm?
The device requires approximately 9 seconds to complete the rhythm analysis.
What is the energy sequence for the defibrillation shocks?
The SY-C025 follows an energy sequence of 150J, 150J, and 200J using a Biphasic Truncated Exponential output.
Does the device provide guidance during use?
Yes, it features 22 audible prompts and LED visual prompts to guide the user through the rescue process.
How fast can the device be ready for a shock?
It takes less than 30 seconds from the initiation of rhythm analysis to readiness for discharge with a new battery.
What are the operating temperature limits?
The AED is designed to operate effectively in temperatures ranging from 0°C to 50°C.
Is there a safety limit for patient impedance?
Yes, the output is automatically disabled if the patient impedance is outside the range of 20Ω to 200Ω.